Manager:
Yinghua Chen, PhD
Contact Email: yinghua.chen@case.edu
Contact Phone: 216-368-8962
Availablility: By Appointment
|
Major Applications:
SPR device measuring binding specificity, affinity and kinetics between molecules via label-free and real-time technique;
Measure ACTIVE concentration of biomolecules;
Supports the use of 96 and 384 well microplates;
For a typical analysis of 7 minutes per cycle, up to 384 samples per run may be processed during 48 hours of unattended operation
Major specifications:
Analysis per cycle: 2-15min
Automation: 48 hours unattended operation
Sample type: low MW drug candidates to high molecular weight proteins
Flow rate range: 1-100 µL/min
Sample compartment: temperature-controlled, 4-45 °C
Flow cell volume: 0.06 µL
Flow cell height: 40 µm
Injection volume: 2-350 µL
Association rate constant (On-rates, Ka): 103 to 107 M-1 S-1 (and higher for macromolecular analytes)
Dissociation rate constant (Off-rates, Kd): 10-5 to 0.5S-1
Molecular weight detection: >100 Da
Baseline noise: < 0.1 RU
Baseline drift: < 0.3 RU/min
Sample refractive index range: 1.33-1.39
For information on theory of Biacore T100, please visit: http://www.youtube.com/watch?
Biacore T200: | $90.00 per day |
Consumable: | $ |